TempTale® Ultra

  • 最高支援六個溫度警報設定 
    Supports up to six temperature alarms
  • 全方面驗證為生命科學市場
    Fully validated for the Life Sciences market
  • 不被IATA’s Dangerous Goods Regulations 強制要求貼危險品標籤
    Does not require labeling per IATA’s Dangerous Goods Regulations
  • 適合生命科學產業/一般產業
TempTale® Ultra Specifications
Monitoring Recording OptionsSingle Use, Single-Use with User Programmable Overwrite, or Multiple Use
Temperature Measurement Range-30°C to +70°C
Temperature Accuracy Range
  • ± 0.5°C from -10°C to 4°C;
  • ± 1.0°C from 45°C to 70°C;
  • ± 1.0°C from -3°C to -10°C
Temperature Resolution0.1°C/F over full temperature measurement range
Data Memory TypeNon-volatile 16K Flash memory
Data Storage Capacity16,000 data points (16K)
Battery Life / Type
  • 12 months operating life / 3.0v coin cell lithium battery – non-replaceable;
  • Multi-Use: Two (2) 3.6v coin cell lithium batteries
Data Sampling IntervalProgrammable from 30 seconds up to a maximum of 120 minutes
Water Resistant HousingIP64 (with protective cap over USB connector)
Start Up Delay0 minutes to 24 hours
Programmable Alarms
  • Programmable high and low limits -> alarm is triggered when temperature exceeds set limits;
  • Up to six alarm types (single temperature limit, temperature range, dual temperature range, MKT, duration)
Alarm Delay Time ThresholdsSingle event (continuous) or cumulative
LCD Indicators
  • Started, start-up delay, running, stopped; Alarm status: OK (✓), Alarmed (✗);
  • Trip summary data (average, minimum, maximum); Current temperature reading;
  • Time remaining / time elapsed against alarm-time thresholds;
  • Marked event
Start Up OptionsManual push button or automatic launch option
Dimensions
  • 108mm L x 50.2mm W x 12.5mm H;
  • Multi-Use: 108mm L x 50.2mm W x 15.2mm H
Weight51g (1.8 oz); Multi-Use: 65g (2.3 oz.)
Quality Assurance Certifications
  • CE, UKCA, RTCA – D016ØG compliant, RoHS compliant, Response time tested to U.S. Pharmacopeia (USP) <1118> and EN 12830 standards;
  • 3-point NIST® Traceable Certificate of Calibration;
  • Validation Manual
Minimum System RequirementsWindows® PC w/ Windows® 7 or higher; Adobe® Reader 9.0 or higher
Data Analysis and Cold Chain Continuous Improvement SoftwareTempTale Manager® Desktop 8.0 or higher; ColdStream® Select
Computer InterfaceUSB 2.0, A-type plug
Ship air, ground, ocean in accordance with –
  • IATA Dangerous Goods Regulations;
  • UN 3091, Packing Instruction 970 Section II – Excepted Lithium Metal and Lithium Alloy Cells and Batteries;
  • U.S. Department of Transportation Lithium Battery Hazardous Material Regulations 49 CFR Parts 171-180;
  • International Maritime Dangerous Goods Regulations, Packing Instruction 903