| Monitoring Recording Options | Single Use, Single Use with User Programmable Overwrite |
| Measurement Range | -95°C to +30°C (-139°F to +86°F) |
| Temperature Accuracy Range |
- ±1.0°C from -90°C to +30°C (±1.8°F from -130°F to +86°F)
- ±1.5°C from -95°C to -90°C (±2.7°F from -139°F to -130°F)
|
| Temperature Resolution | 0.1°C over full temperature measurement range |
| Data Storage Capacity | 16,000 data points (16K) |
| Operating Life / Battery Type |
- 30 days of continuous exposure to dry ice temperatures at or below -80°C
- 5 days of continuous exposure at -95°C
- Two (2) TL-2450 3.6v coin cell lithium batteries—non-replaceable
|
| Data Sampling Interval | Programmable from 30 seconds up to a maximum of 15 minutes in 30 second increments |
| Start Up Delay | 0 minutes to 24 hours in 30 second increments |
| Programmable Alarms |
- Programmable high and low limits à alarm is triggered when temperature exceeds set limits
- Up to six alarms (single temperature limit, temperature range, dual temperature range, MKT, duration)
|
| Alarm Delay Time Thresholds | Single event (continuous), cumulative |
| LCD Indicators | - Started, start-up delay, running, stopped
- Alarm status: OK (✓), Alarmed (✗)
- Trip summary data (average, minimum, maximum)
- Current temperature reading
- Time remaining / time elapsed against alarm thresholds
- Total Time Above High/Low Alarm limits
|
| LED Indicators |
- Blinking Startup LED
- Blinking Stop LED
- File generation in process
- Printer-ready mode activated for printing of PDF report to USB printer
|
| Start Up Options | Manual push button or automatic launch option |
| Typical Dimensions | 108mm L x 50.2mm W x 15.2mm H |
| Weight | Approximately 65 g (2.3 oz) |
| PDF Languages | English, German, French, Spanish, Chinese |
| PDF File Formats | PDF or PDF/A |
| IP Rating | IP64 (with silicon cap on) |
| Time Accuracy | ± 5 minutes per month when exposed to temperatures at or above -90°C |
| Quality Assurance Certifications | - CE, RTCA – D016ØG compliant, RoHS compliant, Response time tested to U.S. Pharmacopeia (USP) <1118> and EN 12830 standards
- 3-point NIST® Traceable Certificate of Calibration
- Validation Manual
|
| Data Analysis Software | TempTale® Manager Desktop Software 8.3 or higher ColdStream® Cold Chain Manager Software 1.8B or higher |
| Minimum System Requirements | Windows PC w/ Windows 7 or higher; Adobe Reader 9.0 or higher |
| Computer Interface | USB 2.0, A-type plug |
| Ship Air in Accordance With | IATA Dangerous Goods Regulations |