Monitoring Recording Options | Single Use, Single Use Overwrite, Multi Use |
Temperature Measurement Range | -30°C to +70°C (-22°F to +158°F) |
Temperature Accuracy Range |
- ±0.5°C from -10°C to +45°C (±0.9°F from +14°F to +113°F)
- ±1.0°C from -30°C to -10°C (±1.8°F from -22°F to +14°F)
- ±1.0°C from +45°C to +70°C (±1.8°F from +113°F to +158°F)
|
Temperature Resolution | 0.1ºC/F over full temperature measurement range |
Humidity Measurement Range | 10% – 90% RH, non-condensing |
Humidity Accuracy Range |
- ±3.0% RH from 20% RH – 80% RH
- ±5.0% RH from 80% RH – 90% RH
- ±5.0% RH from 10% RH – 20% RH
|
Relative Humidity Resolution | 0.1% RH |
Data Storage Capacity | 32,000 data points (16,000 temperature /16,000 humidity) |
Operating Life / Battery Type | 12 months operating life / 3.6v coin cell lithium battery – non-replaceable |
Data Sampling Interval | Programmable from 30 seconds up to a maximum of 120 minutes in 30 second increments |
Start Up Delay | 0 minutes to 24 hours in 30 second increments |
Programmable Alarms |
- High and Low Temperature alarms
- High and Low Humidity alarms
|
Alarm Delay Time Thresholds | Single event (continuous), cumulative |
LCD Indicators |
- Status: Started, Start-up Delay, Running, Stopped
- Alarm status: OK (✓), Alarmed (✗);
- Trip summary data (average, minimum, maximum)
- Current temperature and humidity readings
|
Start Up Options | Manual push button or automatic launch option |
Dimensions | 108mm L x 50.2mm W x 15.2mm H |
Weight | Approximately 60 g (2.1 oz) |
PDF Languages | English, German, French, Spanish, Chinese |
PDF File Formats | PDF or PDF/A |
IP Rating | IP54 (with silicon cap on) |
Time Accuracy | ± 2 minutes per month on average at temperatures (0°C to +50°C) |
Data Analysis Software | TempTale® Manager Desktop 8.3 or higher ColdStream® Cold Chain Manager 1.8B or higher |
Minimum System Requirements | Windows PC w/ Windows 7 or higher; Adobe Reader 9.0 or higher |
Computer Interface | USB 2.0, A-type plug |
Quality Assurance Certifications |
- CE, RTCA – D016ØG compliant, RoHS compliant,
Response time tested to U.S. Pharmacopeia (USP) <1118> and EN 12830 standards
- 3-point NIST® Traceable Certificate of Calibration
- Validation Manual
|
Ship Air in Accordance With | 2018 IATA Dangerous Goods Regulations 59th Edition |